4

Pressure Cheat Sheet

Localize in 10 seconds
  • Pdrain pre-pump, usually negative during active drainage. Reflects how hard pump is pulling. Set limits for the circuit and patient using its IFU.
  • Ppre upstream of oxygenator — after pump.
  • Ppost downstream of oxygenator — closest to patient on return side.
  • ΔP = Ppre − Ppost trend reflects oxygenator resistance. Compare at similar blood flow. A rising gradient can suggest obstruction.
Increasingly negative drainage pressure + low flow Reduce RPM to break suction, then assess volume, bleeding, obstruction and cannula position. Plasma-free hemoglobin supports hemolysis assessment. LDH alone does not establish suction injury.
High return pressure at comparable blood flow
Downstream obstruction — check cannula kink, patient position, clamps. Risk of tubing rupture / circuit decoupling rises with sustained high pressure.
Pattern recognition — 4 grids
A
Both ↑ together
↑Ppre
↑Ppost
Downstream obstruction. Arterial-line kink, cannula clamp, patient malposition, peripheral cannula against vessel wall. Look first. Don't ↑ RPM.
B
Both ↓ together
↓Ppre
↓Ppost
Upstream / preload issue. Hypovolemia, venous cannula kink, tip suction, RA collapse. ↑ RPM worsens it. Volume, position, sedation.
C
ΔP rising at comparable flow
↑Ppre
→Ppost
Suspect oxygenator obstruction / clot. Assess gas exchange and hemolysis with ECMO/perfusion. A rising gradient reflects resistance, compared at similar blood flow.
D
Drain ↓↓ + chugging
↓↓Pdrain
~flow
Suction event. Audible chugging, oscillating flow. Immediately ↓ RPM. Then: volume, position, image cannula. Treat hemolysis if prolonged.
5

Low-Flow Differential

Afterload · Preload · Obstruction
A

Afterload-limited

SVR / MAP too high
  • Clue circuit pressures rise · MAP very high · flow falls at given RPM. Common just after vasopressor up-titration.
  • Mechanism centrifugal pump is afterload-sensitive — pump cut-off curve flattens flow as resistance rises.
  • Fix reduce / hold pressors · vasodilate if appropriate (nicardipine, NTG) → then ↑ RPM. Individualize MAP to perfusion and LV loading.
B

Preload / drainage

Chugging — most common
  • Clue flow oscillates · audible chugging · drain pressure swings very negative · pressures drop when flow drops.
  • Causes hypovolemia, bleeding, tamponade, pneumothorax, excessive RPM, cannula tip against wall / septum.
  • First move ↓ RPM to stable flow before anything else.
  • Then assess fluid responsiveness and bleeding, echo, patient position and cannula. Give volume when indicated, then re-titrate slowly.
C

Mechanical obstruction

Persistent, not oscillating
  • Clue low flow steady (not oscillating) · drain pressure stays very negative · no volume response.
  • Causes kink, thrombus in circuit, tip migration, cannula too shallow (iliac/IVC), DVT propagation, abdominal compartment syndrome.
  • Workup tubing walk-through · CXR / fluoro · echo (cannula tip) · bladder pressure if abdomen distended.
  • Fix reposition · OR re-do if confirmed kink / thrombus · escalate to perfusion + CT surgery.
6

Oxygenator Function & When to Change

ΔP · gas exchange · hemolysis
Daily checks
  • Post-oxygenator saturation and PaO2 assess membrane output, not native lung or upper-body oxygenation.
  • Post-oxygenator ABG: compare at similar blood flow, FdO2 and pre-membrane saturation. Sensor saturation alone can mislead.
  • Gas source confirm blender / wall vs tank during transports — silent failure.
  • Hemolysis / consumption plasma-free hemoglobin, LDH, platelets, fibrinogen and circuit findings. Test frequency follows stability.
"Sigh / burp" technique
Gas-side condensate can ↓ transfer. ECMO/perfusion: brief ↑ sweep only per oxygenator IFU, then restore prescribed sweep. Monitor gases and avoid hypocapnia. Repeat intermittently only if permitted. Recheck post-oxy gases at baseline settings.
When to plan a change-out
  • Trending PaO2 falling on post-oxy ABG at comparable flow, FdO2 and inlet saturation.
  • ΔP rising with clinical impact (hemolysis, oxygenation, alarms).
  • Hemolysis rising LDH, falling haptoglobin, brown urine.
  • DIC pattern fibrinogen falling, d-dimer rising, plt consumption.
  • Visible clot in oxygenator on inspection (flashlight).
Plan, don't crash Plan with perfusion + CT surgery for a controlled exchange. Use trends and patient impact to plan an exchange before catastrophic failure. No single age or PaO2 threshold mandates exchange.
Differential — gas exchange failure
  • Gas source off / disconnected → check blender, tank pressure, line connection.
  • CO2 clearance inadequate → assess gas supply and sweep. Increasing sweep chiefly changes CO2 removal.
  • Oxygenator condensation → device-permitted sigh + reassess (P.7).
  • Membrane clot / failure → gas exchange worsens, with or without a ΔP rise → assess for exchange.
  • Recirculation / shunt on V-V or V-AV configurations — re-evaluate cannula tips.
  • Differential hypoxia on peripheral VA — see P.4.
7

Air / Bubble Management

Prevent · detect · respond
High-risk moments
  • Opening a central line to atmosphere near drainage (IJ tip in SVC/RA).
  • Loose connections during transport or line manipulation.
  • Aggressive flushing or contrast near bubble sensor.
  • Drainage cannula tip suctioning intermittently exposed to air (chugging + open lines = disaster).
Rule No open-to-air manipulation of any line that drains into SVC/RA while on ECMO. Pause infusions, cap, then resume.
If air alarms
  • 1. Look — pump head, oxygenator, tubing for visible air.
  • 2. If air in arterial limb → Stop pump · clamp · 100% FiO2 · perfusion STAT.
  • 3. Venous / pump air → ECMO/perfusion isolates and de-airs the affected segment using the circuit procedure. Do not push air forward.
  • 4. Reset bubble detector only when circuit confirmed clear.
False positives
  • Contrast microbubbles (Definity, Optison) can trigger bubble alarm / pump stop.
  • If ultrasound contrast is needed, agree on a device-specific plan with perfusion before injection. Do not bypass an air safety response solely because contrast was given.
  • Keep infusion and sampling connections closed and air-free. Follow the circuit-specific sampling procedure.
Stroke is the cost
Arterial air on VA → high stroke risk. Treat every bubble alarm as real until proven otherwise.
Readiness — biochemical
Lactate clearing · end-organ recovery · adequate perfusion pressure on low vasoactive support.
Readiness — pulsatile
Aortic valve opens with sustained pulsatility and stable perfusion during support reduction.
Readiness — echo
Assess both ventricles, AV opening, LVOT VTI and filling pressures at reduced support. Thresholds below are cohort clues.
Do not wean
Active bleed · circuit issue · rising lactate · arrhythmia storm · refractory hypoxia · severe lung disease. Fix first.
14

Weaning & Turn-Down Trial

Staged · controlled · echo-driven
Readiness domains — assess together
  • Etiology corrected — the cause of cardiac failure has been treated (PCI, ablation, anti-inflammatory, time).
  • Pulsatile waveform — sustained ejection, interpreted with AV opening and ventricular loading.
  • MAP holds — > 65 without high-dose inopressors.
  • Metabolic resolved — lactate normal · pH normal · K corrected.
  • Lungs / AC — lung function not severely impaired · tolerating anticoagulation.
Turn-down trial workflow
  • Stabilize first — bleeding / circuit issues addressed. Timing follows the cause and evidence of recovery.
  • Stepwise flow ↓ — use the team's controlled protocol with continuous pressure/echo assessment and serial perfusion checks. Stop deterioration promptly.
  • Minimum flow trial — the minimum and duration depend on circuit and program technique. Confirm adequate anticoagulation and rescue readiness before low-flow testing.
  • Echo focus — LV recovery (LVOT VTI, EF, AV opening) + RV function (TAPSE, S', size). RV often easier on TTE than TEE.
  • Pass → plan controlled percutaneous or surgical decannulation with vascular closure and rescue support ready.
  • Fail → back to support, reassess destination on next MDT round.
High vasoactive need argues against removal
Persistent high vasoactive requirements or deteriorating perfusion during turn-down argue against removal. Restore support and reassess.
Staged decannulation strategies
Standard
Pass turn-down → percutaneous or surgical decannulation in OR · vascular closure (Perclose / suture / cut-down) · daily limb exam.
Chronic ↓ EF
If LV dysfunction persists but RV and lung function permit, consider transition to isolated LV support. ECMO can worsen LV afterload, so good systemic perfusion does not establish LV recovery.
RV failure
Consider an appropriate RV-support configuration when isolated RV failure persists. Account for pulmonary vascular resistance, oxygenation and device suitability.
Bi-V failure
Begin LVAD/transplant evaluation early if recovery is unlikely. Timing follows etiology, neurologic and end-organ trajectory, candidacy and goals.
Non-recovery
Destination conversation: LVAD candidacy, transplant listing, or transition to comfort. Engage palliative early.
Limb & access after decann Continue access-site and distal-limb surveillance after removal. Use the program ultrasound strategy to detect thrombosis, pseudoaneurysm or arterial injury. Reassess urgently for swelling, pain, ischemia or bleeding.
▸

Weaning framework — what each trial tests

V-A and V-V differ
V-A — wean the flow
  • Support reduction tests native cardiac output, filling pressures and biventricular reserve. Assess systemic arterial gas exchange as flow falls.
  • Echo clues LVOT VTI ≥10 cm, LVEF >20–25% and lateral mitral S′ ≥6 cm/s were associated with success in a small observational cohort, not validated stand-alone cutoffs.
  • Technique flow reduction, a brief clamped trial or pump-controlled retrograde trial off require a trained team and a defined circuit protocol. Low-flow support can still mask RV failure.
  • Stop for falling perfusion pressure, rising filling pressures, ventricular dilation, hypoxemia or intolerance. Restore support and reassess the cause.
V-V — keep flow, wean gas
  • Ready native lungs can maintain acceptable oxygenation and ventilation on a lung-protective ventilator strategy.
  • Wean gas reduce FdO₂ and sweep using the VV protocol, then assess a monitored sweep-off trial.
  • Maintain blood flow sufficient to limit circuit stasis during the gas-off trial. Minimum flow follows the circuit and protocol.
Other temporary devices (IABP · Impella · TandemHeart) & detailed device-by-device decannulation technique → MCS Weaning & Decannulation card.
15

Exit pathways & destination

Set at cannulation · revisit daily
Bridge to…
Recovery
Myocarditis · post-MI stunning · post-cardiotomy · refractory VT/VF. Turn-down → decannulate. Expect days, not hours.
VAD
Chronic ↓ EF. Impella 5.5 overlap → decannulate ECMO → HF ICU → LVAD eval. Tx-listed pts often go direct.
Transplant
Listed candidate or rapid eval feasible. Requires acceptable oncologic / functional candidacy + center pathway.
Palliation
No reversible cause + no destination — engage palliative early, name surrogate, document plan in EHR.
TBD
Trajectory unclear → daily MDT, family meeting in 24 h, set time-limited trial with explicit milestones and a reassessment date.
When to call who
Perfusion
Immediately for critical alarms, suspected air, loss of flow or oxygenator failure. Involve the team early for planned exchange.
CT surgery
Cannulation · decannulation · bleeding from access · suspected dissection · cannula migration / rupture.
MCS limb team
Limb ischemia · pseudoaneurysm · DPC placement / failure · compartment syndrome.
Advanced HF
LVAD / Tx evaluation · destination decisions · LV venting strategy.
Palliative
Family meetings · goals-of-care conflict · time-limited trial framing · re-cannulation discussions.
Three immediate circuit threats (1) Air in the arterial limb. (2) Bleeding + over-anticoagulation. (3) Decannulation / cannula rupture by movement or migration. Hard-stop checks for each.
VitalFlow™
Centrifugal · touchscreen · MC3 / Medtronic
Illustrative control layout
NAUTILUS DOCK 120 MIN P IN 270 mmHg P Δ 22 mmHg P OUT 248 mmHg V 3.82 L / min SO₂ IN 73.2 % SO₂ OUT 99.9 % XX:XX MM/DD/YYYY T OUT 36.5 °C 2000 RPM VitalFlow™ 3 2 1 0 RPM ×1000 1 2 3 4 5 6 Stylized — refer to Medtronic IFU for exact layout
1Oxygenator dock sits on top. Clip-on as single piece
2Status light bar green/yellow/red — read from across the room
3Tile screen all pressures, sats, temp on one glance
4Big numbers Flow (L/min) + RPM — the two you titrate
5Three hard keys silence · alarm settings · lock
6Physical knob RPM — confirm response and delivered flow
First touch
Confirm flow, alarms and lock state after every change. Use the exact console version's instructions.
Numbers on screen
Pin · PΔ · Pout · SO2 in/out · T · Flow · RPM all integrated. Confirm SaO2 with a real ABG daily — sensors drift.
If screen freezes
Confirm actual flow and activate the trained emergency support procedure. A separate emergency handcrank is available. Do not assume any control remains functional during a console fault.
Don't
Not for cardiotomy suction. Centrifugal pump — needs preload. Don't ↑ RPM to fix chugging.
Cardiohelp
Compact · HLS clip-on · Getinge / Maquet
Transport / legacy
STATUS MAQUET FLOW 2.11 L/min RPM 2150 P pre 152 P post 177 SvO₂ 71% Hb 11.2 CARDIOHELP HLS SET PUMP HEAD A B C D E F Stylized — refer to Getinge IFU for exact layout
AHLS Set — pump + oxygenator as one disposable, clips to right
BTile screen + tap-to-confirm hard keys ✓ ✕ alongside
CVenous probe values Hb · HCT · SvO2 · Temp on the line
DPump well HLS pump head clips directly into console
ERetaining bar verify secure attachment before transfer
FStatus LEDs on the bar — green / yellow / red, visible from across room
VitalFlow
One disposable
HLS Set combines pump head + oxygenator. Change-out is one swap, not two. Don't bump the cassette during transfers.
Venous probe
Live Hb · HCT · SvO2 · Temp on screen — available values depend on the installed sensors and console version. Values drift over days — confirm with lab daily.
Emergency
Know the emergency drive location and rehearse transfer with perfusion. Verify motor engagement and forward flow during recovery.
When you'll see it
US labeling revised January 2025 describes specified surgical bypass uses for <6 h. Prolonged ECMO is outside that label. Verify the current regional IFU and safety notices.
B

Console Buttonology — VitalFlow vs Cardiohelp

High-yield only
 
Medtronic VitalFlow
Maquet Cardiohelp
Lock / unlock
Identify lock state and RPM control on the installed software. Confirm flow and re-lock after changes.
Use the installed version's unlock and confirmation sequence. Confirm displayed flow after changes.
Set low-flow alarm
Individualize the low-flow threshold and verify the alarm response.
Individualize the threshold and review after flow targets change.
Bubble / air action
Visual + audible alarm. Stop pump = arterial air action when configured.
Verify the active bubble response. Contrast requires a planned, supervised strategy rather than automatic safety bypass.
Pressures / sats
Pre + post + ΔP + pre/post sat all on the tile screen — trend in chart.
Pre + post + ΔP shown. Integrated probe adds Hb · HCT · SvO2 · Temp — confirm with ABG.
Emergency mode
Use the emergency support procedure for the fault. Locate and inspect the separate handcrank.
Know the emergency drive and transfer procedure. Confirm actual forward flow, not motor rotation alone.
Sweep / FdO₂
Blender on top of oxygenator. Verify gas source on transport.
Verify the separate gas supply, blender, sweep and backup cylinder. Console power does not prove gas delivery.
C

Cannula Quick Reference

Adult VA · verify with IFU + your team
Role Common product Sizes (Fr) Notes · positioning
V Drainage — fem-RA Bio-Medicus Life Support™ multi-stageMedtronic 21 · 23 · 25 Multi-stage with fenestrations distributed along shaft · tip at cavoatrial junction · TEE/fluoro confirm · mark depth at skin.
V-V — right-atrial dual-lumen Crescent™ RA dual-lumenMedtronic 13 · 15 · 19 Pediatric-sized V-V catheter with both drainage and reinfusion lumens. Not an adult VA single-stage drainage cannula or a generic IJ return adjunct.
V-V — bicaval dual-lumen Crescent™ jugular bicavalMedtronic 24 · 26 · 28 · 30 · 32 For long-term V-V (not VA). RIJ → bicaval. Confirm position by TEE / fluoro. Recirculation if malpositioned.
A Return — femoral artery Bio-Medicus arterial / Opus™ access kitMedtronic 15 · 17 · 19 Match vessel caliber and pressure–flow needs. Larger cannulae may compromise limb perfusion. Plan and verify distal perfusion.
A Distal perfusion (DPC) Antegrade SFA sheath (institutional) 5 · 6 · 7 Plan antegrade SFA perfusion at femoral VA cannulation. Use a secure, de-aired circuit connection and verify actual distal flow.
S

Six scenarios — the ones the console will throw at you

VitalFlow sim · bedside
01 Low flow + chugging Preload
Flow2.1
RPM3400
P drain−110
P pre120
SO₂ in72%
Audiblechug
Stop
↓ RPM first — break suction before anything else. Never ↑ RPM to "fix" low flow.
Think
Tip suction · hypovolemia · bleeding · tamponade · pneumothorax · cannula against wall.
Do
Check bleeding and fluid responsiveness, echo, position and cannula. Give volume only when indicated, then re-titrate.
Call
Call ECMO/perfusion early, immediately if perfusion is impaired. Involve the cannulating team for suspected migration.
02 Condensation: post-oxy sats fall Gas side
SO₂ out↓
PaO₂ out↓
ΔPstable
Flowstable
Clot seennone
Exhaustwet
Check
Confirm post-oxy gases at unchanged flow / FdO₂. Check gas supply, connections and exhaust patency.
Think
Gas-side condensation can impair exchange. No visible clot or ΔP rise does not exclude thrombosis or membrane failure.
Sigh
ECMO/perfusion: brief gas-side ↑ sweep per oxygenator IFU, if gases permit → restore prescribed sweep. Monitor CO₂ and stop for hypocapnia. Repeat intermittently only if IFU / protocol permit.
Recheck
Recheck gases at baseline. Persistent decline or wet exhaust → urgent assessment for other causes, including plasma leak, and need for exchange.
03 Bubble alarm — arterial air Air
BubbleDETECT
PumpSTOPPED
Flow0.0
RPM0
SpO₂ R rad88%
Ptdeteriorating
Stop
stop pump · clamp arterial · 100% FiO2.
Think
Treat an air alarm as real until the circuit is checked. Otherwise air source = open central line near drainage, loose connection, suction event with open line.
Do
De-air via designed ports — do not push air forward. Reset bubble detector only when circuit confirmed clear.
Call
Perfusion STAT. Document neuro exam. Treat stroke pathway if focal deficit.
04 North–South syndrome Differential
Flow4.5
SO₂ out99%
R rad SpO₂82%
Fem SpO₂98%
PEEP8
AV opensyes
Stop
Don't be fooled by femoral sat. Right-radial sampling screens the upper body, but does not directly measure coronary oxygenation.
Think
Native LV ejecting desaturated blood to arch / coronaries / brain while ECMO returns to lower body.
Do
Optimize lungs and reassess flow. Discuss V-AV or alternate arterial return for persistent upper-body hypoxemia. Unload for LV distension or stasis, not as a fixed step for every case.
Call
ECMO MD + CT surgery early if V-AV / central conversion likely.
05 Arrest on VA ECMO Pseudo-PEA
A-lineflat
RhythmVF / asys
Flow4.2
MAP62
Pulsatilitylost
SO₂ out99%
Stop
Assess effective perfusion immediately. If ECMO maintains adequate pressure and organ perfusion, loss of pulsatility alone is not an indication for compressions. If circulation is inadequate or uncertain, activate resuscitation and restore support.
Think
Circuit first: flow OK? Pump running? Decannulation / rupture? Air? Pump-head clot? If circuit fails → ALS until ECMO re-established.
Do
Treat the rhythm and reversible cause with the resuscitation/ECMO team. Use current defibrillator-specific ALS dosing when arrest care is needed. Set ventilation from lung function and gases, with the appropriate resuscitation or ECMO ventilation strategy.
Call
ICU consultant + ECMO coordinator immediately. Remember the pt may be awake — sedate/analgese before painful interventions.
06 Cannula rupture / decannulation Exsanguination
Sitebleeding
Flow↓↓
MAP42
Hb↓
P drain−0 / +
Air alarmpossible
Stop
Clamp circuit proximal & distal to the breach · turn pump off. Direct pressure to the site. Never re-suture a leaking cannula — needle puncture ruptures it.
Think
Return (arterial) breach → rapid exsanguination + air risk. Drainage breach → air entrainment. Cause: suture needle, withdrawal, disconnection, cut-down site bleed.
Do
Massive transfusion · pressure · ALS if no flow. Recannulate (rewire if possible, otherwise new site) — OR or cut-down. Resume ECMO once circuit intact.
Call
CT surgery STAT · perfusion · ICU attending · blood bank — activate MTP. Secure airway if not already.
Always
comes
first
1 · Patient
Look at the patient before the console. Pulse · perfusion · pupils · vent. The screen is data. The patient is the truth.
2 · Reduce RPM
For suction/chugging, reduce RPM to break suction. Other low-flow causes require rapid circuit and patient assessment before speed changes.
3 · Call early
Perfusion + ECMO MD. Two minds beat one at 3 am. "I'm not sure" is the right thing to say on the phone.
▶

VA-ECMO learning resources

For residents · fellows · attendings
Console simulators
  • VitalFlow™ training sim browser-based — walk the tile dashboard, alarms, and emergency mode.
    infusemeddev.com/projects/medtronic/vitalflow_sim
  • Cardiohelp / HLS Getinge training portal — touch flow, HLS module orientation.
    getinge.training
Complications & troubleshooting
  • Low-flow troubleshooting intensiveblog.com/oh-no-low-flow
  • Venous insufficiency / chugging intensiveblog.com/an-ecmo-earthquake
  • Differential hypoxia (N–S) intensiveblog.com/different-hypoxia-va-ecmo
  • Cardiac arrest on VA ECMO intensiveblog.com/cardiac-arrest-on-ecmo-does-it-still-count
  • Cannula rupture / dislodgement intensiveblog.com/everything-ecmo-018
Curricula & guidelines
  • ELSO General Guidelines for All ECLS Cases · Adult Cardiac Failure Guideline · ECMO Specialist curriculum. elso.org
  • EuroELSO Lorusso/EACTS expert consensus on V-A ECMO (2021).
  • INTENSIVE / Alfred ICU Everything ECMO series (Nickson, Hoffman, Diehl, Burrell) — case-based Q&A.
  • CICU teaching structured CICU rounds curriculum (Gottula & Elliott).
A therapeutic assay does not guarantee a safe circuit. Interpret anticoagulation alongside bleeding, circuit clot, hemolysis and cardiac stasis. Use a validated program nomogram with a stated agent, assay, target and recheck interval. The exact aPTT window varies by reagent and patient.
16

Choose the agent for the patient

UFH or direct thrombin inhibitor
AgentWhy choose it?What changes the plan?
Unfractionated heparinEstablished ECMO agent, familiar assays and protamine reversal. Requires antithrombin.Bleeding, HIT, antithrombin deficiency and assay discordance.
BivalirudinDirectly inhibits thrombin without antithrombin. An option for HIT or a program DTI strategy.Renal dysfunction, changing clearance and stagnant blood. No specific antidote. ECMO dosing is off-label and differs from PCI dosing.
ArgatrobanAnother non-heparin option, including when renal dysfunction favors it.Hepatic dysfunction can markedly prolong effect. Use an agent-specific protocol.
Evidence boundary: adult comparisons favoring bivalirudin are predominantly observational and cannot establish a survival advantage. A small pediatric pilot randomized trial does not establish superiority in adult VA-ECMO.
17

Match the measurement to the drug

trend and reconcile discordance
MeasurementUseful forCommon trap
Anti-XaUFH activity using the laboratory's heparin-calibrated method.Not a measure of bivalirudin. Hemolysis, bilirubin, lipemia and assay design can alter results.
aPTTCommonly used for UFH or DTI titration, with an agent-specific target.Baseline coagulopathy, lupus anticoagulant and acute-phase changes can disconnect the value from drug effect.
ACTA rapid whole-blood adjunct where included in a validated strategy.Affected by platelets, temperature, anemia and factor levels. Do not equate it with a drug concentration.
Dilute thrombin time / ecarinDrug-calibrated DTI assessment when available and aPTT is unreliable.Availability and calibration differ. Consult the coagulation laboratory.
Recheck after change
  • Check after initiation, dose adjustment, bleeding, organ-function change or CRRT change. Stable monitoring frequency must be explicit.
  • With heparin resistance, examine assay validity and antithrombin before escalating blindly.
Watch patient and circuit
  • Trend CBC, fibrinogen, coagulation tests, visible clot, ΔP at comparable flow and membrane gas transfer.
  • Use plasma-free hemoglobin when available for hemolysis. LDH and haptoglobin are supporting, nonspecific markers.
18

Bleeding, procedures and clearance

source control first
  • Major bleeding identify and control the source, reduce or interrupt anticoagulation with the ECMO team, and replace demonstrated deficits. Reassess when it is safe to restart.
  • Bivalirudin label half-life is about 25 min with normal renal function and about 3.5 h in dialysis patients. These are pharmacokinetic observations, not guaranteed procedure-hold times.
  • CRRT can change clearance in either clinical direction as renal function, filter and prescription change. Recheck the assay rather than automatically raising the dose.
  • Transfusion individualize hemoglobin, platelet and fibrinogen goals to bleeding, procedures and oxygen delivery. Active bleeding generally calls for at least as much hemostatic support as the nonbleeding state.
  • Acquired vWS shear-related loss of large multimers can contribute to mucosal/GI bleeding despite near-normal routine coagulation tests.
  • HIT assess probability and test appropriately. Stop all heparin exposure when clinically indicated and use an appropriate non-heparin anticoagulant.
19

Viscoelastic testing complements the workup

interpret the complete tracing
FindingInterpretation in a bleeding patient
Prolonged R / CTAnticoagulant effect or factor deficiency. Drug-related prolongation alone is not an indication for plasma.
Low angle or clot strengthUse platelet count and fibrinogen-specific testing to distinguish the deficits before targeted replacement.
Excess clot lysisConfirm hyperfibrinolysis and the clinical context. Antifibrinolytics require a deliberate benefit–thrombosis assessment.
Stasis remains dangerous: a non-ejecting LV, aortic root and poorly flowing side branches can thrombose despite a therapeutic systemic assay. No minimum circuit-flow number guarantees safety during an anticoagulation hold. Check the chamber and the circuit, then reassess anticoagulation and unloading together.
ECMO team members STOP & complete the checklist together — shift change (RN–RN) · after cannulation · before & after mobilization · arrival post-travel (RN / Perfusion / Provider). Mobilize with a trained ECMO team and an agreed plan. This check targets decannulation, air entry and power/gas failure.
19

The STOP checklist

every item, out loud
Cannulas & connections
  • ☐ Arterial + venous cannula location & site assessment
  • ☐ Cannulas secured in two places. All connections checked
  • ☐ Tubing without kinks / sharp turns. Pressure lines flushed
Power · gas · emergency-ready
  • ☐ Individualized low-flow alert and safety actions verified · emergency power confirmed
  • ☐ Trace oxygen/air supply by labels and connectors · full backup gas · emergency drive
  • ☐ Backup equipment ready under the program plan · clamps and approved connectors · emergency clamp site identified
☐ Team contacts & action plan at the bedside.  ☐ Backup readiness includes circuit, gas, power and emergency drive availability. Priming/storage limits follow the component IFU and program policy. For power/driver failure, call immediately and follow the practiced procedure to prevent backflow and restore forward flow.
▤

Cannula securement — the three peripheral configurations

schematic — not to scale
FEMORAL VENOUS · drainage tip → cavo-atrial jct mark depth at skin 2 stay sutures + device + dressing → drainage to pump FEMORAL ARTERIAL + DPC return → iliac/aorta 2 sutures + device DPC → distal leg AXILLARY (side-graft) 8 mm Dacron side-graft allows ambulation · watch arm hyperperfusion antegrade → arm
20

Securement standards & site pearls

two points · marked depth · the limb is the risk
SiteSecure — every cannula: 2-point fixation + marked depthWatch
All cannulasUse device-compatible fixation at two independent points. Mark insertion depth, anchor tubing to prevent traction and maintain a clean visible dressing.Recheck fixation and depth at every handoff and move. Stop movement if unsafe and obtain immediate team help. Sedation does not substitute for securement.
Femoral venousLong multi-stage drainage cannula, tip at the cavo-atrial junction.Drainage obstruction, malposition and air entrainment. Classic return-to-drain recirculation applies to venous return in VV/V-AV, not simple femoral VA proximity.
Femoral arterialAntegrade distal-perfusion cannula (DPC). Suture introducer at hub + eyelet, loop tubing (no kinks), Tegaderm so site + loop visible.Limb ischemia is the headline risk — distal Doppler / pulses. Harlequin.
Axillary (graft)End-to-side 8 mm Dacron side-graft → antegrade flow & ambulation.Arm hyperperfusion / compartment syndrome.
21

Perfusion touchbase & the ECMO record

daily + q-hourly
Team touchbase — every day and change
  • Mode & cannulation? Flow steady? Any bleeding? Which anticoagulant, assay and target?
  • Clots / ΔP across the oxygenator? SvO₂ & Hct last updated? Weaning or road trips today (perfusion must travel)?
What perfusion charts (q-hourly + per change)
  • Flow/RPM/sweep/FiO₂ · ABP/CVP/PAP/Pven/pre-post-oxy pressures · SpO₂/SvO₂/Hct/temp/ACT.
  • Record support times, configuration, cannula sites/sizes/depth, gas settings, anticoagulant, assays, alarms and the current destination.
The lifeline rule: two-point fixation, a marked & documented depth, and tubing anchored to the patient — verified every shift & before every move (proning, transport, CT). Chart depth · sutures/device · dressing · distal pulses · flows/pressures. Check for bleeding, infection, ischemia, thrombosis and migration / dislodgement.